GE Healthcare, Xeleris 2 Processing & Review Workstation. The display, processing, archiving, and communication of data acquired by Emission Tomography cameras used in diagnostic radiology, including procedures for planar imaging, whole body imaging, tomographic (SPECT) imaging, positron imaging by coincidence, attenuation correction, and anatomical image registration
Published May 1, 2013 · Recall initiated: October 6, 2011
Summary
FDA records medical device Class II recall Z-1164-2013, published May 1, 2013 and initiated October 6, 2011. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 11; Lot: Mfg Lot or Serial # 0000CZC7105FJM000 00000CZC827NRN000 00000CZC8283V4000 0000CZC8283V47000 0000CZC8283V72001 0000CZC8283TX5001 00000000211423001 0000CZC8161XJT000 0000CZC94154WV003 0000CZC83942HL000 0000CZC8396CC9001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE Healthcare has become aware of a software issue in your Xeleris 2.1 workstation (that is used for processing Discovery 530/570 data), that may impact patient safety. On imported CT scans, Myocardial Perfusion Imaging (MPI) studies may create artifacts in attenuation corrected images. This may lead to a potential for misdiagnosis.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution - USA including GA, and NY; and countries of: Canada, Israel, Italy and Switzerland.
- Lot
- Mfg Lot or Serial # 0000CZC7105FJM000 00000CZC827NRN000 00000CZC8283V4000 0000CZC8283V47000 0000CZC8283V72001 0000CZC8283TX5001 00000000211423001 0000CZC8161XJT000 0000CZC94154WV003 0000CZC83942HL000 0000CZC8396CC9001
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDAVarian brand RPM Respiratory Gating System, RPM Gating v1.7.5 with 3D Option, Model Number: HS2 Product Usage:May 1, 2013
- FDANeuViz 16 Multi-Slice CT Scanner System, Part number (PN): 989605858501, Model: NeuViz 16. This is a computedMay 1, 2013
Questions about this recall
Was GE Healthcare, Xeleris 2 Processing & Review Workstation. The display, processing, archiving, and communication of data acquired by Emission Tomography cameras used in diagnostic radiology, including procedures for planar imaging, whole body imaging, tomographic (SPECT) imaging, positron imaging by coincidence, attenuation correction, and anatomical image registration recalled?
Yes. FDA published recall Z-1164-2013 on May 1, 2013. Reported hazard: GE Healthcare has become aware of a software issue in your Xeleris 2.1 workstation (that is used for processing Discovery 530/570 data), that may impact patient safety. On imported CT scans, Myocardial Perfusion Imaging (MPI) studies may create artifacts in attenuation corrected images. This may lead to a potential for misdiagnosis.
What should I do if I have this product?
Voluntary: Firm initiated · Letter