Back to search
FDAFDA · Food and drugs·Recall number Z-1175-2013

ORTHOCORD One Violet and One Blue Braided Composite Suture with MO-6 Taper Needles, Product Code: 223115. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.

Published May 1, 2013 · Recall initiated: March 22, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
DePuy Mitek, Inc., a Johnson & Johnson Co.
Distributor
Nationwide Distribution.
Lot
ALL from Jan 1 2010 - Corrected IFU implementation date
Risk classification
Class II
Area (FDA)
device
Country
United States