FDAFDA · Food and drugs·Recall number Z-1183-2013
Diacap(R) Ultra Dialysis Fluid Filter, Catalog Number 7107367. Intended to filter bacteria and endotoxins from dialysis fluid.
Published May 8, 2013 · Recall initiated: March 1, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is the potential of the residual moisture of the Diacap Ultra membrane being reduced over time, which results in the reduction of the membrane permeability (ultrafiltration coefficient).
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- B. Braun Medical, Inc.
- Distributor
- Distribution and manufacturing dates for the expansion lots was reported as 5/23-6/14/2013 and 1/22-24/2013 respectively. Nationwide Distribution including the states of AZ, CA, DE, FL, GA, HI, IA, ID, IL, KS, KY, LA, MN, MO, NJ, NM, NY, OH, OR, PA, SC, SD, TN, TX, UT, WA and WV.
- Lot
- Lot numbers 302950410, 309910111, 313050411, 314300511, 316040711, 316450711, 305880812, 306301012, and 306311012. This recall was expanded on/about June 18, 2013 to include additional lot #s 305010113 and 305000113.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States