Back to search
FDAFDA · Food and drugs·Recall number Z-1191-2017

LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.

Published February 22, 2017 · Recall initiated: January 10, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Encore Medical, Lp
Distributor
US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI
Lot
1005528A, 1006882A, 1107305A, 1101798A, 1100565A, 1005531A, 1211716A, 1007408A, 1004038B, 1005533D, 1106805A, 1203471A, 1005533C, 1100565A, 1101795A, 1007079A, 1005532A, 1100560A
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card