NxStage Cartridge Express with Speedswap, Catalog No. CAR-535. An accessory device for facilitating the removal and replacement of a flow-compromised filter during treatment of acute and chronic renal failure or fluid overload.

Published March 6, 2024 · Recall initiated: January 29, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Product is recalled due to reports of blood leaking or spraying from the arterial line on the cartridge.If this failure occurs, a patient may experience uncontrolled blood loss or a healthcare professional could be exposed to infectious patient blood.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
NxStage MDS Corporation
Distributor
US domestic distribution to Dialysis Centers located in AZ, CA, CO, GA, IL, IN, MA, MD, MI, MN, MO, MS, NC, NJ, NY, TN, TX, VA, WI. No international distribution.
Lot
UDI-DI M535CAR5350; Package DI M535CAR5355 Lots 20878001, 30978013, 30978032, 31178011, 31178019
Risk classification
Class II