FDAFDA · Food and drugs·Recall number Z-1195-2018

Nucleic Acid test intended for screening for MRSA/SA. Each kit has 20 tubes. in vitro diagnostic.

Published April 4, 2018 · Recall initiated: October 27, 2017

Summary

FDA records medical device Class II recall Z-1195-2018, published April 4, 2018 and initiated October 27, 2017. Recalling firm: Roche Molecular Systems, Inc.; Status: Terminated; Reported quantity: 136 kits; Lot: Lot #s: 70712A (6 kits); 70601W (76 kits); 70713A (4 kits); 70718A (50 kits); and 70809A (kits were not distributed to end users). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The cobas MRSA/SA tests may have decreased performance compared to the analytical sensitivity for MSRA detection. A screening failure can lead to potential infection in the colonized patient being tested as well as the spread of MSRA to others. The United States is not impacted, as the cobas MRSA/SA nucleic acid test for use on the cobas liat system is not available in the US.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Roche Molecular Systems, Inc.
Distributor
Worldwide distribution. Austria, Denmark, Germany, Singapore, Netherlands, Hong Kong, and Switzerland. US product is not affected.
Lot
Lot #s: 70712A (6 kits); 70601W (76 kits); 70713A (4 kits); 70718A (50 kits); and 70809A (kits were not distributed to end users).
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Nucleic Acid test intended for screening for MRSA/SA. Each kit has 20 tubes. in vitro diagnostic. recalled?

Yes. FDA published recall Z-1195-2018 on April 4, 2018. Reported hazard: The cobas MRSA/SA tests may have decreased performance compared to the analytical sensitivity for MSRA detection. A screening failure can lead to potential infection in the colonized patient being tested as well as the spread of MSRA to others. The United States is not impacted, as the cobas MRSA/SA nucleic acid test for use on the cobas liat system is not available in the US.

What should I do if I have this product?

Voluntary: Firm initiated · Letter