REF 623401 R, S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Right, Medium, Sterile R Qty 1 The S-ROM Hinged knee femoral component is used in revision surgeries when there is significant bone loss or ligament instability and additional support is needed. S-Rom femorals and inserts are brought into nearly every revision case as backup should the MCL become compromised during the procedure. If additional constraint is required the surgeon can use the S-ROM femoral and insert. S-ROM femorals are used in less than 10% of knee revision procedures, but are present almost all the time.
Published March 19, 2014 · Recall initiated: February 24, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM¿ Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the p
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- DePuy Orthopaedics, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, Australia, Austria, China, Czech Republic, Denmark, Dubai / UAE, Estonia, France, Germany, Hungary, Iberia, India, Ireland, Israel, Italy, Malaysia, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Russia, Singapore, Slovenia, Spain, South Africa, Sweden, Switzerland, Turkey, United Kingdom, and Vietnam.
- Lot
- 102325, 102326, 109315, 109316, 120082, 128475, 142843, 142845, 162487, 197812, 197813, 197814, 207924, 222654, 222657, 231689, 231690, 236435, 236442, 261758, 265711, 265714, 265715, 265717, 282997, 285154, 285155, 289096, 302804, 302807, 302808, 317754, 317755, 338154, 338155, 338157, 354580, 360635, 360636, 366432, 378645, 378647, 378650, 400505, 400506, 402456, 414735, 414736, 414737, 430481, 451998, 451999, 452448, 456741, 456742, 456743, 472193, 472194, 472195, 197401RA, D2WBJ4, DY5EF4, E3SD14, E49HN4, EW6HW4, EW6HX4, and FK8CR4
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States