FDAFDA · Food and drugs·Recall number Z-1207-2019

OnSight 3D Extremity System- X-Ray, Tomography Computed System Catalog Numbers: 1743566 (w/ Attached Console) 1743574 (w/ Remote Console) The device is intended to be used for x-ray computed tomography and projection x-ray imaging of upper and lower extremities of adult patients and pediatric patients aged 12 and over.

Published May 1, 2019 · Recall initiated: March 19, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When the user performs the re-assignment of a parent / companion pair, the parent volume is transferred to the new patient but the companion volume will remain in the original patient s exam.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Carestream Health, Inc.
Distributor
Worldwide Distribution: US (nationwide) to states of: Arizona, California, New York, Alabama, Georgia, Georgia, Illinois, Minnesota, New York, Oregon, Pennsylvania; and countries of: IRAN, ITALY, JERUSALEM, KOREA, PHILIPPINES, PORTUGAL,SAUDI ARABIA, SINGAPORE, SWITZERLAND THAILAND, and UNITED KINGDOM.
Lot
US Serial Numbers: 58008585 58007506 58000020 58008493 58000015 58007502 58007504 58008494 58007503 58008018 58007505 58008500 58008011 58008014 OUS: 10012 20005 20012 20016 20007 20017 10018 10004 10007 20004 20008 10023 10009 20019 20013 10002 10003 10014 10010 20006 20022 20003 20010
Risk classification
Class II
Area (FDA)
device
Country
United States