OnSight 3D Extremity System- X-Ray, Tomography Computed System Catalog Numbers: 1743566 (w/ Attached Console) 1743574 (w/ Remote Console) The device is intended to be used for x-ray computed tomography and projection x-ray imaging of upper and lower extremities of adult patients and pediatric patients aged 12 and over.
Published May 1, 2019 · Recall initiated: March 19, 2019
Summary
FDA records medical device Class II recall Z-1207-2019, published May 1, 2019 and initiated March 19, 2019. Recalling firm: Carestream Health, Inc.; Status: Terminated; Reported quantity: 37 units; Lot: US Serial Numbers: 58008585 58007506 58000020 58008493 58000015 58007502 58007504 58008494 58007503 58008018 58007505 58008500 58008011 58008014 OUS: 10012 20005 20012 20016 20007 20017 10018 10004 10007 20004 20008 10023 10009 20019 20013 10002 10003 10014 10010 20006 20022 20003 20010. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When the user performs the re-assignment of a parent / companion pair, the parent volume is transferred to the new patient but the companion volume will remain in the original patient s exam.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Carestream Health, Inc.
- Distributor
- Worldwide Distribution: US (nationwide) to states of: Arizona, California, New York, Alabama, Georgia, Georgia, Illinois, Minnesota, New York, Oregon, Pennsylvania; and countries of: IRAN, ITALY, JERUSALEM, KOREA, PHILIPPINES, PORTUGAL,SAUDI ARABIA, SINGAPORE, SWITZERLAND THAILAND, and UNITED KINGDOM.
- Lot
- US Serial Numbers: 58008585 58007506 58000020 58008493 58000015 58007502 58007504 58008494 58007503 58008018 58007505 58008500 58008011 58008014 OUS: 10012 20005 20012 20016 20007 20017 10018 10004 10007 20004 20008 10023 10009 20019 20013 10002 10003 10014 10010 20006 20022 20003 20010
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADRX Revolution Mobile X-Ray SystemDecember 13, 2023
- FDASiemens Digital Linear Accelerators of type: ARTISTE MV System MEVATRON M2/Primus Mid-Energy PRIMUS PRIMUS HIMay 1, 2019
- FDACovidien Endo Clinch II Auto Suture Grasper, 5mm REF: 174317 The Covidien Endo Clinch" II Auto Suture" grasperMay 1, 2019
- FDACARESTREAM DIRECTVIEW CR Software Generates digital mammographic images that can be used for screening andDecember 24, 2014
- FDAKodak DirectView DR 7500 Diagnostic X-Ray System Product Usage: The DR 7500 system is a permanently installedAugust 20, 2014
- FDAKODAK DirectView DR 7500 Dual Detector System, MODEL DR 7500, Manufactured by Carestream Health, Inc. Made inFebruary 5, 2014
Questions about this recall
Was OnSight 3D Extremity System- X-Ray, Tomography Computed System Catalog Numbers: 1743566 (w/ Attached Console) 1743574 (w/ Remote Console) The device is intended to be used for x-ray computed tomography and projection x-ray imaging of upper and lower extremities of adult patients and pediatric patients aged 12 and over. recalled?
Yes. FDA published recall Z-1207-2019 on May 1, 2019. Reported hazard: When the user performs the re-assignment of a parent / companion pair, the parent volume is transferred to the new patient but the companion volume will remain in the original patient s exam.
What should I do if I have this product?
Voluntary: Firm initiated · Letter