Back to search
FDAFDA · Food and drugs·Recall number Z-1240-2016

Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757 Product Usage: For in vitro diagnostic use in the quantitative determination of estradiol in human serum and plasma (heparinized and EDTA) using the ADVIA Centaur¿, ADVIA Centaur XP, ADVIA Centaur XPT and ADVIA Centaur CP systems.

Published March 30, 2016 · Recall initiated: January 13, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc
Distributor
Worldwide Distribution  US Nationwide , AR, AU, AU, BD, BO, BR, CA, CH, CL, CN, CO, DO, EC, EG, HK, ID, IE, IL, IN, JP, KR, LK, LY, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VE, VN, and ZA.
Lot
All lots
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.