Brivo MR355, NMRI system
Published March 5, 2025 · Recall initiated: January 31, 2025
Summary
FDA records medical device Class II recall Z-1240-2025, published March 5, 2025 and initiated January 31, 2025. Recalling firm: GE Healthcare (China) Co., Ltd.; Status: Ongoing; Reported quantity: 22; Lot: Brivo MR355 (No UDI) System IDs 082427010096 082427020046 082427040126 082427070076 082427070077 082427070170 082427100245 082427120151 082427180030 082427210147 082427210148 082427220025 082427220045 082427230060 082427240060 082427310104 0856270058 0856270059 MRR10200 MRR9920 RU1063MR03 RU8902MR01 (Updated 12/29/25 - Additional System ID 83027805544513) (Updated 1/28/26 - Additional System IDs: 82427220047 82427240055 82427070163 82427060068 82427140137 82427010122 82427280018) (Updated 2/19/26: Additional System ID EG1736MR02). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare (China) Co., Ltd.
- Distributor
- Domestic US distribution nationwide. International distribution worldwide.
- Lot
- Brivo MR355 (No UDI) System IDs 082427010096 082427020046 082427040126 082427070076 082427070077 082427070170 082427100245 082427120151 082427180030 082427210147 082427210148 082427220025 082427220045 082427230060 082427240060 082427310104 0856270058 0856270059 MRR10200 MRR9920 RU1063MR03 RU8902MR01 (Updated 12/29/25 - Additional System ID 83027805544513) (Updated 1/28/26 - Additional System IDs: 82427220047 82427240055 82427070163 82427060068 82427140137 82427010122 82427280018) (Updated 2/19/26: Additional System ID EG1736MR02)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- China
Related recalls
- FDABD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump andMarch 5, 2025
- FDAASP BIOTRACE Auto Read 60 Steam BI Process Challenge Device, REF 73135.March 5, 2025
- FDAGE Healthcare Optima 680 Expert, Model number 6670000; X-ray/computed tomography systemDecember 11, 2024
- FDAGE Healthcare Revolution Eagle, Model Number 5849504-3, X-ray/computed tomography systemDecember 11, 2024
- FDAGE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as:December 11, 2024
- FDAGE Healthcare Revolution ACE, Model number 6670000-3; X-ray/computed tomography systemsDecember 11, 2024
Questions about this recall
Was Brivo MR355, NMRI system recalled?
Yes. FDA published recall Z-1240-2025 on March 5, 2025. Reported hazard: For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
What should I do if I have this product?
Voluntary: Firm initiated · Letter