Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R endoscopes are intended for use in endoscopic diagnosis and treatment within the ureter and kidney.
Published March 8, 2017 · Recall initiated: December 12, 2016
Summary
FDA records medical device Class II recall Z-1252-2017, published March 8, 2017 and initiated December 12, 2016. Recalling firm: Olympus Corporation of the Americas; Status: Terminated; Reported quantity: 1,627 units (769 URF-V2 and 858 URF-V2R); Lot: All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Olympus has received complaints regarding the breakage of the endoscope's insertion tube bending section during surgical procedures. Some of these complaints are associated with tissue trauma, including perforation, and insertion tubes which were stuck inside the patient and had to be surgically removed.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Olympus Corporation of the Americas
- Distributor
- Worldwide distribution-US Nationwide and country of: Canada.
- Lot
- All serial numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOlympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.October 7, 2026
- FDAOlympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.October 7, 2026
- FDAOlympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunctionOctober 7, 2026
- FDAOlympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA.September 30, 2026
- FDAURF-P6 Uretero-reno fiberscope and URF-P6R Uretero-reno fiberscopeMarch 8, 2017
- FDASiemens ADVIA Centaur Systems Calibrator A Kits, Cat No. 04800646, SMN 10285903 (2 pack)Cat No. 04800735, SMNMarch 8, 2017
Questions about this recall
Was Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R endoscopes are intended for use in endoscopic diagnosis and treatment within the ureter and kidney. recalled?
Yes. FDA published recall Z-1252-2017 on March 8, 2017. Reported hazard: Olympus has received complaints regarding the breakage of the endoscope's insertion tube bending section during surgical procedures. Some of these complaints are associated with tissue trauma, including perforation, and insertion tubes which were stuck inside the patient and had to be surgically removed.
What should I do if I have this product?
Voluntary: Firm initiated · Letter