Back to search
FDAFDA · Food and drugs·Recall number Z-1252-2017

Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R endoscopes are intended for use in endoscopic diagnosis and treatment within the ureter and kidney.

Published March 8, 2017 · Recall initiated: December 12, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Olympus has received complaints regarding the breakage of the endoscope's insertion tube bending section during surgical procedures. Some of these complaints are associated with tissue trauma, including perforation, and insertion tubes which were stuck inside the patient and had to be surgically removed.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Olympus Corporation of the Americas
Distributor
Worldwide distribution-US Nationwide and country of: Canada.
Lot
All serial numbers
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card