FDAFDA · Food and drugs·Recall number Z-1252-2017

Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R endoscopes are intended for use in endoscopic diagnosis and treatment within the ureter and kidney.

Published March 8, 2017 · Recall initiated: December 12, 2016

Summary

FDA records medical device Class II recall Z-1252-2017, published March 8, 2017 and initiated December 12, 2016. Recalling firm: Olympus Corporation of the Americas; Status: Terminated; Reported quantity: 1,627 units (769 URF-V2 and 858 URF-V2R); Lot: All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Olympus has received complaints regarding the breakage of the endoscope's insertion tube bending section during surgical procedures. Some of these complaints are associated with tissue trauma, including perforation, and insertion tubes which were stuck inside the patient and had to be surgically removed.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Olympus Corporation of the Americas
Distributor
Worldwide distribution-US Nationwide and country of: Canada.
Lot
All serial numbers
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R endoscopes are intended for use in endoscopic diagnosis and treatment within the ureter and kidney. recalled?

Yes. FDA published recall Z-1252-2017 on March 8, 2017. Reported hazard: Olympus has received complaints regarding the breakage of the endoscope's insertion tube bending section during surgical procedures. Some of these complaints are associated with tissue trauma, including perforation, and insertion tubes which were stuck inside the patient and had to be surgically removed.

What should I do if I have this product?

Voluntary: Firm initiated · Letter