Back to search
FDAFDA · Food and drugs·Recall number Z-1257-2014

Natural-Knee¿ II Posterior Stabilized (P.S.) Condylar Tibial Inserts: N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS INS TIB SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ N-K II CONS PS TIB INS SZ

Published March 26, 2014 · Recall initiated: February 25, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Lot
Item Number 672509100 672509101 672509200 672509203 672511100 672511103 672513101 672513103 672513201 672516103 672519103 672522201 672525103 672525203
Risk classification
Class II
Area (FDA)
device
Country
United States