AXIOM Artis Systems Product usage: Artis zee is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Published April 2, 2014 · Recall initiated: February 7, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Certain units of the AXIOM Artis Systems are using a flat detector cooling unit with inappropriate connectors that may possibly leak, causing a system failure and possible drop out of the gantry system.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
USA Nationwide Distribution in the states of WI, KY, OH, NJ, NC, FL, PA, MA, IA, TX, NY, IL, MI, AZ, CA, and LA.
Lot
Model nos. 7412807, 7727717, 7728392 with serial numbers: 35020, 34999,35098 35113 35042 35061 35064 35073 35014 35069 35053 35055 35056 35054 35008 35057 35044 35018 35097 35108 35046 35048 35050 35045 35004 35019 46004 46005 40000 40009 40011 40006 40007 40010 44016 44005 44011
Risk classification
Class II
Area (FDA)
device
Country
United States