Zilver PTX Drug-Eluting Peripheral Stent, - 7 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery). The Paclitaxel coating helps prevent the artery from narrowing again (restenosis). The Zilver; PTX stent acts as a scaffold to: Hold open a narrowed artery in the thigh (femoropopliteal artery) caused by PAD. Improve blood flow to the extremity.
Published May 22, 2013 · Recall initiated: April 18, 2013
Summary
FDA records medical device Class I recall Z-1275-2013, published May 22, 2013 and initiated April 18, 2013. Recalling firm: Cook, Inc.; Status: Terminated; Reported quantity: 535 units in USA; Lot: REF ZIV6-35-125-7-80-PTX ; UDI 10827002248950 all lots distributed through 4/16/13. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cook, Inc.
- Distributor
- Worldwide Distribution - USA Nationwide; in the countries of Japan, Europe, Brazil, and Australia
- Lot
- REF ZIV6-35-125-7-80-PTX ; UDI 10827002248950 all lots distributed through 4/16/13
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAZilver® PTX® Drug-Eluting Peripheral Stent - 6 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTXMay 22, 2013
- FDAAMS 800® Urinary Control System. The AMS 800® Urinary Control System is used to treat urinary incontinence dueMay 22, 2013
- FDAGrippo's HOT DILL PICKLE Artificially Flavored Potato Chips, 1®oz, 3.5oz, INGREDIENTS: SELECTED POTATOES,May 22, 2013
- FDAAMS 800® Urinary Control System. The AMS 800® Urinary Control System is used to treat urinary incontinence dueMay 22, 2013
- FDALam Sheng Kee Mini Sausage Hong Kong Style, Net Weight 6.3 oz.), Lam Sheng Kee (HK) International Limited, RM3May 22, 2013
- FDAInsight Instruments MIS ( Multi-Port Illumination System) Illuminated pilot tubes work like a ceiling lightMay 22, 2013
Questions about this recall
Was Zilver PTX Drug-Eluting Peripheral Stent, - 7 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery). The Paclitaxel coating helps prevent the artery from narrowing again (restenosis). The Zilver; PTX stent acts as a scaffold to: Hold open a narrowed artery in the thigh (femoropopliteal artery) caused by PAD. Improve blood flow to the extremity. recalled?
Yes. FDA published recall Z-1275-2013 on May 22, 2013. Reported hazard: Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation
What should I do if I have this product?
Voluntary: Firm initiated · Letter