FDAFDA · Food and drugs·Recall number Z-1275-2013
Zilver PTX Drug-Eluting Peripheral Stent, - 7 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery). The Paclitaxel coating helps prevent the artery from narrowing again (restenosis). The Zilver; PTX stent acts as a scaffold to: Hold open a narrowed artery in the thigh (femoropopliteal artery) caused by PAD. Improve blood flow to the extremity.
Published May 22, 2013 · Recall initiated: April 18, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Cook, Inc.
- Distributor
- Worldwide Distribution - USA Nationwide; in the countries of Japan, Europe, Brazil, and Australia
- Lot
- REF ZIV6-35-125-7-80-PTX ; UDI 10827002248950 all lots distributed through 4/16/13
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States