Transonic Hemodialysis Monitor, Model HD03 -- Product Usage: The Hemodialysis Monitor is intended for use by trained medical personnel to measure delivered blood flow, vascular access recirculation, vascular access blood flow and cardiac output on patients receiving hemodialysis treatment.

Published March 25, 2015 · Recall initiated: February 17, 2015

Summary

FDA records medical device Class II recall Z-1295-2015, published March 25, 2015 and initiated February 17, 2015. Recalling firm: Transonic Systems Inc; Status: Terminated; Reported quantity: Domestic: 21 units, Foreign: 10 units; Lot: Serial Numbers HD03A60004; HD03A81167; HD03A81225; HD03A81237; HD03A91338; HD03A91394; HD03A91406; HD03B01461; HD03B21782; HD03B21803; HD03A60020; HD03A71082; HD03A81207; HD03A91255; HD03A91316; HD03Bl1679; HD03B11684; HD03B11705; HD03B21812; HD03B31971; HD03A81240; HD03A71086; HD03B01459; HD03B01594; HD03B11667; HD03B42175; HD03B42160; HD03B42161; HD03A91258; HD03A70046; HD03B31901. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Customer unable to obtain measurements due to the device seeing no flow or irrelevant flow condition during dialysis procedures. The user saw a zero or a very low flow value displayed on HD03 and/or no measurements produced in other modes (recirculation; access flow; or cardiac output).

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Transonic Systems Inc
Distributor
Worldwide Distribution - USA Nationwide and the countries of: Australia, Canada, and South Korea
Lot
Serial Numbers HD03A60004; HD03A81167; HD03A81225; HD03A81237; HD03A91338; HD03A91394; HD03A91406; HD03B01461; HD03B21782; HD03B21803; HD03A60020; HD03A71082; HD03A81207; HD03A91255; HD03A91316; HD03Bl1679; HD03B11684; HD03B11705; HD03B21812; HD03B31971; HD03A81240; HD03A71086; HD03B01459; HD03B01594; HD03B11667; HD03B42175; HD03B42160; HD03B42161; HD03A91258; HD03A70046; HD03B31901
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Transonic Hemodialysis Monitor, Model HD03 -- Product Usage: The Hemodialysis Monitor is intended for use by trained medical personnel to measure delivered blood flow, vascular access recirculation, vascular access blood flow and cardiac output on patients receiving hemodialysis treatment. recalled?

Yes. FDA published recall Z-1295-2015 on March 25, 2015. Reported hazard: Customer unable to obtain measurements due to the device seeing no flow or irrelevant flow condition during dialysis procedures. The user saw a zero or a very low flow value displayed on HD03 and/or no measurements produced in other modes (recirculation; access flow; or cardiac output).

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit