FDAFDA · Food and drugs·Recall number Z-1304-2024

DCA Systems HbA1c Reagent Kit

Published March 27, 2024 · Recall initiated: February 7, 2024

Summary

FDA records medical device Class II recall Z-1304-2024, published March 27, 2024 and initiated February 7, 2024. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Ongoing; Reported quantity: 74,211 Kits (Expanded 2/28/24: additional 72,892 kits); Lot: Siemens Material Number: 10311134; UDI/DI: 00630414532806; Lot#: 9640093, 0645103, 0648103, 0649103, 0654103 . Expanded Lots (02/28/2024): 0637093, 0641093, 0642093, 0647103, 0653103, 0655103. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Worldwide distribution - US Nationwide and the country of Taiwan.
Lot
Siemens Material Number: 10311134; UDI/DI: 00630414532806; Lot#: 9640093, 0645103, 0648103, 0649103, 0654103 . Expanded Lots (02/28/2024): 0637093, 0641093, 0642093, 0647103, 0653103, 0655103;
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was DCA Systems HbA1c Reagent Kit recalled?

Yes. FDA published recall Z-1304-2024 on March 27, 2024. Reported hazard: The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.

What should I do if I have this product?

Voluntary: Firm initiated · Letter