DCA Systems HbA1c Reagent Kit
Published March 27, 2024 · Recall initiated: February 7, 2024
Summary
FDA records medical device Class II recall Z-1304-2024, published March 27, 2024 and initiated February 7, 2024. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Ongoing; Reported quantity: 74,211 Kits (Expanded 2/28/24: additional 72,892 kits); Lot: Siemens Material Number: 10311134; UDI/DI: 00630414532806; Lot#: 9640093, 0645103, 0648103, 0649103, 0654103 . Expanded Lots (02/28/2024): 0637093, 0641093, 0642093, 0647103, 0653103, 0655103. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the country of Taiwan.
- Lot
- Siemens Material Number: 10311134; UDI/DI: 00630414532806; Lot#: 9640093, 0645103, 0648103, 0649103, 0654103 . Expanded Lots (02/28/2024): 0637093, 0641093, 0642093, 0647103, 0653103, 0655103;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;October 7, 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mLAugust 26, 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mLAugust 26, 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULAugust 12, 2026
- FDAAccess Total T4, REF 33800, IVD, thyroxine reagent assay.March 27, 2024
- FDATENSIVE Conductive Adhesive Gel, REF 016-401600March 27, 2024
Questions about this recall
Was DCA Systems HbA1c Reagent Kit recalled?
Yes. FDA published recall Z-1304-2024 on March 27, 2024. Reported hazard: The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.
What should I do if I have this product?
Voluntary: Firm initiated · Letter