HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgery

Published March 27, 2024 · Recall initiated: March 11, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Baxter Healthcare Corporation
Distributor
US Nationwide distribution in the state of Florida.
Lot
123510000293, 123510000294, 123510000295, 123510000296, 123510000298, 123510000299, 123510000300, 123510000764, 123510000765, 123510000766, 123510000767, 123510000768, 123510000769, 123510000770, 123510000771, 123510000772, 123510000773, 123510000800, 123510000801, 123510000802, 123510000803, 123510000804, 123510000805, 123510000806, 123510000807, 123510000808, 123510000809, 123510000297, 123350001033, 123350001034, 123350001035, 123350001036, 123350001040, 123350001041, 123350001042, 123360000294, 123360000295, 123360000297, 123360000298, 123360000299, 123360000300, 123360000301, 123360000302, 123360000303, 123360000304, 123360000733, 123360000734, 123360000735, 123360000736, 123360000737, 123360000738, 123360000739, 123360000740, 123360001972, 123360001973, 123360001974, 123360001975, 123360001976, 123360001977, 123360001978, 123360001979, 123360001980, 123360001981, 123350000968, 123350000969, 123350000970, 123350000971, 123350000972, 123350000973, 123350000974, 123350000975, 123350000987, 123350000988, 123350000989, 123350000990, 123350000991, 123350000992, 123350000993, 123350000994, 123350001032
Risk classification
Class II
Area (FDA)
device
Country
United States