Bard(R) Premature Infant Feeding tube, ENFit(TM), REF EN0036430 Product Usage: BARD(R) Nasogastric Sump Tubes with ENFit(TM) connector are intended to be used for: Decompression of stomach by suction or aspiration of gastric contents Short-term administration of term tube feeding, lavage fluid and medications Feeding Tubes with the ENFit(TM) Connector are made of a smooth plastic with a two-eyed tube used for nasogastric feeding of neonates, infants and adults as well as diagnostic and therapeutic aspiration. These products are sold single-use, sterile. The Feeding Tube with the ENFit(TM) connector provides a way to reduce the risk of enteral device misconnections. The ENFit(TM) connector: Addresses patient side connections between feeding tubes, administration sets, medication, flush and bolus feeding syringes, and other enteral devices. Does not allow connectivity with any other connector for any other clinical use.
Published April 18, 2018 · Recall initiated: February 27, 2018
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
BMD has identified that the product code/lot number combination identified above may have ineffectively fused tubing and ENFit(TM) connector.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- C.R. Bard, Inc.
- Distributor
- US Nationwide Distribution
- Lot
- Lot Numbers: NGBN2862, NGAXX217, NGAVX225
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States