Back to search
FDAFDA · Food and drugs·Recall number Z-1367-2013

SLIDERPREP PLUS SLIDE PREPARATION CENTRIFUGE. Produce a monolayer of cells onto a glass slide from any fluid suspension.

Published May 29, 2013 · Recall initiated: March 13, 2013

Summary

FDA records medical device Class II recall Z-1367-2013, published May 29, 2013 and initiated March 13, 2013. Recalling firm: Separation Technology, Inc.; Status: Terminated; Reported quantity: 102; Lot: Serial numbers: 0110, 0111, 0112, 0210, 0211, 0212, 0310, 0311, 0312, 0410, 0411, 0412, 0511, 0512, 0611, 0710, 0711, 0811, 0812, 0910, 0911, 0912, 1011, 1012, 1112, 1210, 1212. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Separation Technology, Inc. recalled SlidePrep Plus Cytology Centrifuges manufactured prior to January 1, 2013 for having the potential to start up unexpectedly when encountering an electrostatic discharge.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Separation Technology, Inc.
Distributor
Worldwide Distribution-USA (nationwide) including the states of PA, NC, IL, KS, and FL, and the countries Canada and Turkey.
Lot
Serial numbers: 0110, 0111, 0112, 0210, 0211, 0212, 0310, 0311, 0312, 0410, 0411, 0412, 0511, 0512, 0611, 0710, 0711, 0811, 0812, 0910, 0911, 0912, 1011, 1012, 1112, 1210, 1212.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SLIDERPREP PLUS SLIDE PREPARATION CENTRIFUGE. Produce a monolayer of cells onto a glass slide from any fluid suspension. recalled?

Yes. FDA published recall Z-1367-2013 on May 29, 2013. Reported hazard: Separation Technology, Inc. recalled SlidePrep Plus Cytology Centrifuges manufactured prior to January 1, 2013 for having the potential to start up unexpectedly when encountering an electrostatic discharge.

What should I do if I have this product?

Voluntary: Firm initiated · Letter