SLIDERPREP PLUS SLIDE PREPARATION CENTRIFUGE. Produce a monolayer of cells onto a glass slide from any fluid suspension.
Published May 29, 2013 · Recall initiated: March 13, 2013
Summary
FDA records medical device Class II recall Z-1367-2013, published May 29, 2013 and initiated March 13, 2013. Recalling firm: Separation Technology, Inc.; Status: Terminated; Reported quantity: 102; Lot: Serial numbers: 0110, 0111, 0112, 0210, 0211, 0212, 0310, 0311, 0312, 0410, 0411, 0412, 0511, 0512, 0611, 0710, 0711, 0811, 0812, 0910, 0911, 0912, 1011, 1012, 1112, 1210, 1212. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Separation Technology, Inc. recalled SlidePrep Plus Cytology Centrifuges manufactured prior to January 1, 2013 for having the potential to start up unexpectedly when encountering an electrostatic discharge.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Separation Technology, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including the states of PA, NC, IL, KS, and FL, and the countries Canada and Turkey.
- Lot
- Serial numbers: 0110, 0111, 0112, 0210, 0211, 0212, 0310, 0311, 0312, 0410, 0411, 0412, 0511, 0512, 0611, 0710, 0711, 0811, 0812, 0910, 0911, 0912, 1011, 1012, 1112, 1210, 1212.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABimix Inj 30 :1 Syr, 12x1.5 ml syr, 10x1.2 ml syr, 2x2 ml syr, 2x1 ml syr, 12x.05 ml syr Penile Injection,30June 5, 2013
- FDAPinnacle Hip Solutions ALTRX Polyethylene Acetabular Liner, Ref 1221-36-452 Product Usage: An acetabular linerMay 29, 2013
- FDACobas® 8000 modular analyzer series software version 03-01, 02, 03 and 04, impacting the cobas e 602 moduleMay 29, 2013
- FDAMEDITECH Bi-Directional interactive Communication Analyzer Interface; Integrates with MEDITECH's LaboratoryMay 29, 2013
- FDAStryker Orthopaedics Triathlon Femoral Distal Augment- Left; 1 single unit per package Howmedica OsteonicsMay 29, 2013
- FDAProgressive Medical PTFE Coated Disposable Electrosurgical Electrodes - Push Button Pencil With PTFE Blade andMay 29, 2013
Questions about this recall
Was SLIDERPREP PLUS SLIDE PREPARATION CENTRIFUGE. Produce a monolayer of cells onto a glass slide from any fluid suspension. recalled?
Yes. FDA published recall Z-1367-2013 on May 29, 2013. Reported hazard: Separation Technology, Inc. recalled SlidePrep Plus Cytology Centrifuges manufactured prior to January 1, 2013 for having the potential to start up unexpectedly when encountering an electrostatic discharge.
What should I do if I have this product?
Voluntary: Firm initiated · Letter