VITROS Performance Verifier II
Published April 3, 2024 · Recall initiated: February 27, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics
- Distributor
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
- Lot
- Product Code: 8231474; GTIN: 10758750004577; Lot No. R1176.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States