FDAFDA · Food and drugs·Recall number Z-1380-2022
Tibial Component Modular. orthopedic prosthesis. Model Nos: Small, W 16-2817/32 Medium, W 16-2817/35 Large, W 16-2817/37
Published July 20, 2022 · Recall initiated: May 24, 2022
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Distributor
- Domestic distribution to AL, CA, FL, GA, IL, IN, KS, MI, MN, NV, NY, OH, TN, TX, WA. Foreign distribution to Algeria Argentina Australia Austria Belgium Bulgaria Canada China Colombia Denmark Ecuador Estonia Finland France Germany Great Britain Greece Hungary India Indonesia Israel Italy Latvia Libya Lithuania Mexico Netherlands Norway Pakistan Peru Philippines Poland Romania Saudi Arabia Slovakia Slovenia Spain Sudan Sweden Switzerland Thailand Turkey Ukraine United Arab Emirates Uruguy Vietnam
- Lot
- UDI-DI: 04026575370870 Small, W; 04026575370887 Medium, W; 04026575370894 Large, W; Serial/Lot Numbers: 210308/0020 210308/0035 210222/0789 151116/4741
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany