SYNTHECEL Dura Repair, 510K # K131792, K113071, Class II, Classification name: Dura Substitute Device listing number: D140596, D211012 SYNTHECEL Dura Repair is intended for use as a dura replacement for the repair of dura mater in adults.
Published April 20, 2016 · Recall initiated: December 8, 2015
Summary
FDA records medical device Class II recall Z-1385-2016, published April 20, 2016 and initiated December 8, 2015. Recalling firm: Synthes (USA) Products LLC; Status: Terminated; Reported quantity: 289 units; Lot: SYNTHECEL Dura Repair 5.0 cm x 5.0 cm (2x2), part numbers: SC.400.025.01S, Lot Numbers: 7840495, 7855339, 7860566, 7880950, 7900039, 7909573, 7922342, 7932416, 7984107. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was reported that the SYNTHECEL Dura Repair could adhere to various anatomical structures post implantation, which may expose patients to potential risks in the event a surgeon intends to temporarily apply and subsequently remove a SYNTHECEL implant. SYNTHECEL Dura Repair labeling and promotional material are being updated to clarify that SYNTHECEL is intended to remain in place and is not intended to be removed.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes (USA) Products LLC
- Distributor
- Nationwide: NY, PA, AL, WI, FL, MO, TN, OH, MI, MN, CO, CA, NV, AZ, LA, OR, TX, IA, IN, MS, IL, AR, WV, HI, VA, MD, WA, OK, MA, KY, NJ, CT, No foreign accounts.
- Lot
- SYNTHECEL Dura Repair 5.0 cm x 5.0 cm (2x2), part numbers: SC.400.025.01S, Lot Numbers: 7840495, 7855339, 7860566, 7880950, 7900039, 7909573, 7922342, 7932416, 7984107
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASTARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screwJune 5, 2024
- FDA13 mm Titanium Hindfoot Arthrodesis Cannulated Nails EX, sterile -Intended to facilitate tibiotalocalcanealMarch 13, 2024
- FDARAPIDSORB IPS Battery Pack, Sterile-Intended for use in non-load bearing fracture repair and reconstructiveMarch 13, 2024
- FDAOpal, Intervertebral Cage, 10x32mm, h 11mm, Revolvable-Indicated for degenerative spine disease Part Number:March 13, 2024
- FDAEPIQ DIAGNOSTIC ULTRASOUND SYSTEM, Model EPIQ 5C, EPIC 5G, EPIQ 5W, EPIQ 7C, EPIC 7GC, and EPIQ 7W. DiagnosticApril 20, 2016
- FDACygnus Medical First StepBedside ADDWATER Kit,non-sterile. First Step in cleaning flexible endoscopes. SKU#April 20, 2016
Questions about this recall
Was SYNTHECEL Dura Repair, 510K # K131792, K113071, Class II, Classification name: Dura Substitute Device listing number: D140596, D211012 SYNTHECEL Dura Repair is intended for use as a dura replacement for the repair of dura mater in adults. recalled?
Yes. FDA published recall Z-1385-2016 on April 20, 2016. Reported hazard: It was reported that the SYNTHECEL Dura Repair could adhere to various anatomical structures post implantation, which may expose patients to potential risks in the event a surgeon intends to temporarily apply and subsequently remove a SYNTHECEL implant. SYNTHECEL Dura Repair labeling and promotional material are being updated to clarify that SYNTHECEL is intended to remain in place and is not intended to be removed.
What should I do if I have this product?
Voluntary: Firm initiated · Letter