AURIGA 30 BRAZIL ZERO COST SYSTEM UPN: M0068S30BZ0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.

Published May 29, 2019 · Recall initiated: April 8, 2019

Summary

FDA records medical device Class II recall Z-1396-2019, published May 29, 2019 and initiated April 8, 2019. Recalling firm: Boston Scientific Corporation; Status: Terminated; Reported quantity: 3 (OUS); Lot: Serial Number: 21302966, 21303016, 21303021. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Boston Scientific Corporation
Distributor
Worldwide distribution to Australia, Belgium, Brazil, Chile, Colombia, Egypt, France, Germany, Great Britain, India, Indonesia, Iran, Italy, Lebanon, Malaysia, Mexico, Romania, Russia Fed., South Africa, South Korea, Spain, Taiwan, United Arab Emirates and Vietnam
Lot
Serial Number: 21302966, 21303016, 21303021
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was AURIGA 30 BRAZIL ZERO COST SYSTEM UPN: M0068S30BZ0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery. recalled?

Yes. FDA published recall Z-1396-2019 on May 29, 2019. Reported hazard: Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.

What should I do if I have this product?

Voluntary: Firm initiated · Letter