FDAFDA · Food and drugs·Recall number Z-1396-2019
AURIGA 30 BRAZIL ZERO COST SYSTEM UPN: M0068S30BZ0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.
Published May 29, 2019 · Recall initiated: April 8, 2019
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Boston Scientific Corporation
- Distributor
- Worldwide distribution to Australia, Belgium, Brazil, Chile, Colombia, Egypt, France, Germany, Great Britain, India, Indonesia, Iran, Italy, Lebanon, Malaysia, Mexico, Romania, Russia Fed., South Africa, South Korea, Spain, Taiwan, United Arab Emirates and Vietnam
- Lot
- Serial Number: 21302966, 21303016, 21303021
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States