Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.
Published June 5, 2013 · Recall initiated: April 30, 2013
Summary
FDA records medical device Class II recall Z-1402-2013, published June 5, 2013 and initiated April 30, 2013. Recalling firm: Elekta, Inc.; Status: Terminated; Reported quantity: 31; Lot: 325134, 325254, 325082, 325159, 325268, 325237, 325331, 325267, 325236, 325314, 325127, 325261, 325286, 325077, 325269, 325129, 325296, 325273, 325139, 325201, 325137, 325184, 325161, 325076, 325086, 325112, 325081, 325083, 325099, 325110, 325271. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Elekta, Inc.
- Distributor
- USA Nationwide including the states of CA, FL, IL, MD, MA, NJ, NY, NC, ND, OH, OR, PA, TN, TX, VA, WA, DC and Puerto Rico
- Lot
- 325134, 325254, 325082, 325159, 325268, 325237, 325331, 325267, 325236, 325314, 325127, 325261, 325286, 325077, 325269, 325129, 325296, 325273, 325139, 325201, 325137, 325184, 325161, 325076, 325086, 325112, 325081, 325083, 325099, 325110, 325271
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: Geneva Product Name: Brachytherapy Applicator Product Description: The device is intended forJuly 15, 2026
- FDABrand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: TheJuly 15, 2026
- FDABrand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary Remote Afterloading ApplicatorJuly 15, 2026
- FDAMOSAIQ Oncology Information SystemJune 17, 2026
- FDAAlere" hCG Combo Cassette (20/10 mIU/mL) Kit. Product Usage: The AlereTM hCG Combo (20/10 mIU/mL) test is aJune 5, 2013
- FDAPTS PANELS, Glucose Test Strips for use with CardioChek Brand Analyzers Product Usage: used by healthcareJune 5, 2013
Questions about this recall
Was Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed. recalled?
Yes. FDA published recall Z-1402-2013 on June 5, 2013. Reported hazard: Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.
What should I do if I have this product?
Voluntary: Firm initiated · Letter