PRIMUS HI, Digital Linear Accelerator, Model No. 04504200 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.
Published April 25, 2018 · Recall initiated: February 14, 2018
Summary
FDA records medical device Class II recall Z-1402-2018, published April 25, 2018 and initiated February 14, 2018. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 18; Lot: Serial no: 2958 3832 3241 3574 3519 3922 3924 3361 4036 3302 70-4185 3949 3547 3657 3622 3731 3565 5153. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.
- Lot
- Serial no: 2958 3832 3241 3574 3519 3922 3924 3361 4036 3302 70-4185 3949 3547 3657 3622 3731 3565 5153
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDABrilliance 64 with 4.1.7 XX026 software version model number 728231 Product Usage: Computed Tomography X-rayApril 25, 2018
- FDABrilliance Big Bore (Radiology) - Model no. 728244, Computed Tomography X-Ray system Product Usage: ComputedApril 25, 2018
Questions about this recall
Was PRIMUS HI, Digital Linear Accelerator, Model No. 04504200 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer. recalled?
Yes. FDA published recall Z-1402-2018 on April 25, 2018. Reported hazard: Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit