PRIMUS HI, Digital Linear Accelerator, Model No. 04504200 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

Published April 25, 2018 · Recall initiated: February 14, 2018

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.
Lot
Serial no: 2958 3832 3241 3574 3519 3922 3924 3361 4036 3302 70-4185 3949 3547 3657 3622 3731 3565 5153
Risk classification
Class II
Area (FDA)
device
Country
United States