ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant Part Number: 254400525
Published July 27, 2022 · Recall initiated: June 29, 2022
Summary
FDA records medical device Class II recall Z-1419-2022, published July 27, 2022 and initiated June 29, 2022. Recalling firm: DePuy Orthopaedics, Inc.; Status: Ongoing; Reported quantity: 872 units; Lot: GTIN: 10603295434061 Lots: AB4536235, AB4733296, AB4841493, AB4810808, AB4841495, AB4904565, AB4967214, AB5004557, and AB5034303. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- DePuy Orthopaedics, Inc.
- Distributor
- Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam
- Lot
- GTIN: 10603295434061 Lots: AB4536235, AB4733296, AB4841493, AB4810808, AB4841495, AB4904565, AB4967214, AB5004557, and AB5034303
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAATTUNE AFFIXIUM Cementless Fixed Bearing Knee with 3DP Technology, Size 6. Component of the ATTUNE Total KneeApril 10, 2024
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert SM 12MM, Part Number 151760312March 29, 2023
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert XSM 12MM, Part Number 151760212March 29, 2023
- FDAMODULIFT VBR SZ.M 4MM SET SCREW DRIVER, MF790R, part of Modulift VBR Medium Instrument Set, ST-0441July 27, 2022
- FDASoluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVDJuly 27, 2022
- FDAATTUNE REVISION POSTERIOR FEMORAL AUGMENT SZ 9 12MM-intended as a Component in total knee arthroplasty (TKA)February 16, 2022
Questions about this recall
Was ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant Part Number: 254400525 recalled?
Yes. FDA published recall Z-1419-2022 on July 27, 2022. Reported hazard: Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit