FDAFDA · Food and drugs·Recall number Z-1428-2017
Apollo; Models: 9784000131 9784000231 9784020131 9784020231 9784100231 9784120131 9784120231 9784130131 Product Usage: General, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures.
Published March 15, 2017 · Recall initiated: December 27, 2016
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Villa Sistemi Medicali S.P.A.
- Distributor
- US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, PA, Puerto Rico, TN, TX, UT, and W.I
- Lot
- Serial Numbers: 06100127 06110138 07010149 07020158 07020163 07020165 07030171 07040206 09010566 09090605 10100752 11100873 15041446
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Italy