FDAFDA · Food and drugs·Recall number Z-1430-2022

SIGNA Premier magnetic resonance scanner, model 5748519.

Published July 27, 2022 · Recall initiated: June 24, 2022

Summary

FDA records medical device Class II recall Z-1430-2022, published July 27, 2022 and initiated June 24, 2022. Recalling firm: GE Healthcare, LLC; Status: Ongoing; Reported quantity: 52; Lot: SIGNA Premier GTINs: 00840682135269 and 00195278010797. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Under rare conditions, SIGNA Premier gradient coils could result in elevated acoustic noise during scanning. After prolonged periods of elevated acoustic noise, this could potentially lead to hearing loss.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE Healthcare, LLC
Distributor
Worldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, MI, MN, NY, and WI. There was government distribution but no military distribution. The countries of China, France, Hong Kong, India, Italy, Japan, Korea, Kuwait, Norway, Oman, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Thailand, Turkey, and United Kingdom.
Lot
SIGNA Premier GTINs: 00840682135269 and 00195278010797
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SIGNA Premier magnetic resonance scanner, model 5748519. recalled?

Yes. FDA published recall Z-1430-2022 on July 27, 2022. Reported hazard: Under rare conditions, SIGNA Premier gradient coils could result in elevated acoustic noise during scanning. After prolonged periods of elevated acoustic noise, this could potentially lead to hearing loss.

What should I do if I have this product?

Voluntary: Firm initiated · Letter