FDAFDA · Food and drugs·Recall number Z-1438-2013

adult defibrillation, cardioversion, pacing, monitoring Disposable Electrodes, Model P-211-M1 and P-214-M1. Product Usage: Used by trained professionals in hospitals, doctors offices, and Emergency Medical Services for adult low-energy defibrillation, transcutaneous pacing, cardioversion and monitoring.

Published June 5, 2013 · Recall initiated: April 30, 2013

Summary

FDA records medical device Class II recall Z-1438-2013, published June 5, 2013 and initiated April 30, 2013. Recalling firm: TZ Medical Inc.; Status: Terminated; Reported quantity: US: 10,410 pieces; Outside US: 100 pieces; Lot: Model P-211-M1 Lot numbers: Y081712-17; Y081712-18; Y081712-19; Y081712-20; Y112612-08; Y112612-09; Y112612-10; Y112612-11; Y112612-12; Y112612-13; and Y020113-07. Model P-214-M1 Lot numbers: Y081712-04; Y112612-24; Y112612-23; and Y031913-10. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

TZ Medical Inc is recalling its P-211-M1 and P-214-M1 Medtronic / Physio Control Compatible transparent defibrillation electrodes because it may contain a defective white wire on the connector assembly. If a failure were to occur, the end user would be able to identify the failure by inspecting the white wire 1/4 inch from the Gray Medtronic / Physio adapter. The consignees will notice discolor

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
TZ Medical Inc.
Distributor
Worldwide distribution - USA (nationwide) and Chile.
Lot
Model P-211-M1 Lot numbers: Y081712-17; Y081712-18; Y081712-19; Y081712-20; Y112612-08; Y112612-09; Y112612-10; Y112612-11; Y112612-12; Y112612-13; and Y020113-07. Model P-214-M1 Lot numbers: Y081712-04; Y112612-24; Y112612-23; and Y031913-10.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was adult defibrillation, cardioversion, pacing, monitoring Disposable Electrodes, Model P-211-M1 and P-214-M1. Product Usage: Used by trained professionals in hospitals, doctors offices, and Emergency Medical Services for adult low-energy defibrillation, transcutaneous pacing, cardioversion and monitoring. recalled?

Yes. FDA published recall Z-1438-2013 on June 5, 2013. Reported hazard: TZ Medical Inc is recalling its P-211-M1 and P-214-M1 Medtronic / Physio Control Compatible transparent defibrillation electrodes because it may contain a defective white wire on the connector assembly. If a failure were to occur, the end user would be able to identify the failure by inspecting the white wire 1/4 inch from the Gray Medtronic / Physio adapter. The consignees will notice discolor

What should I do if I have this product?

Voluntary: Firm initiated · Telephone