adult defibrillation, cardioversion, pacing, monitoring Disposable Electrodes, Model P-211-M1 and P-214-M1. Product Usage: Used by trained professionals in hospitals, doctors offices, and Emergency Medical Services for adult low-energy defibrillation, transcutaneous pacing, cardioversion and monitoring.

Published June 5, 2013 · Recall initiated: April 30, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

TZ Medical Inc is recalling its P-211-M1 and P-214-M1 Medtronic / Physio Control Compatible transparent defibrillation electrodes because it may contain a defective white wire on the connector assembly. If a failure were to occur, the end user would be able to identify the failure by inspecting the white wire 1/4 inch from the Gray Medtronic / Physio adapter. The consignees will notice discolor

What to do

Voluntary: Firm initiated · Telephone

Recall details

Manufacturer
TZ Medical Inc.
Distributor
Worldwide distribution - USA (nationwide) and Chile.
Lot
Model P-211-M1 Lot numbers: Y081712-17; Y081712-18; Y081712-19; Y081712-20; Y112612-08; Y112612-09; Y112612-10; Y112612-11; Y112612-12; Y112612-13; and Y020113-07. Model P-214-M1 Lot numbers: Y081712-04; Y112612-24; Y112612-23; and Y031913-10.
Risk classification
Class II
Area (FDA)
device
Country
United States