RespVent Closed Suction System T-Piece Tracheostomy Adult, Diameter: 4.6mm (14F), Length: 30.5cm (12in.), MDI Adapter, Cont. 10 Closed Suction Catheter, Sterile R, RxOnly, Reorder Number: RTG-02300, UDI: 00850108006035 - Product Usage: To be used to aspirate liquids or semi solids from a patient s upper airway.
Published March 11, 2020 · Recall initiated: August 19, 2019
Summary
FDA records medical device Class II recall Z-1443-2020, published March 11, 2020 and initiated August 19, 2019. Recalling firm: RESPIRATORY THERAPEUTICS GROUP LLC; Status: Ongoing; Reported quantity: 723 cases (7230 units); Lot: Lot KS1901013: 20019/03/31 - 06/192019 Lot KS1902002: 2019/06/21 - 07/31/2019. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It has been determined that some units of catheters were not tightly connected to the thumb port. This may result in the catheter disconnecting from the thumb port which could reduce the suction performance and make it difficult to remove from the patient's trachea.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- RESPIRATORY THERAPEUTICS GROUP LLC
- Distributor
- US Nationwide distribution in the states of CA, NJ, NY. OUS: None
- Lot
- Lot KS1901013: 20019/03/31 - 06/192019 Lot KS1902002: 2019/06/21 - 07/31/2019
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASunstar GUM Toothpaste 3 g tubes, Toothpaste made in Japan, packed in China, NDC 53050-0002-01March 18, 2020
- FDACarbidopa and Levodopa Tablets, USP 25 mg/250mg, 100 ct bottle, Rx Only, Mayne Pharma, Greenville, NC 27834.March 18, 2020
- FDAMonitor for the Functional Imaging System for Magnetic Resonance Imaging SystemMarch 11, 2020
- FDAKits containing Covidien Force TriVerse Electrosurgical Devices. Distributed in EMEA (Europe, Middle East, andMarch 11, 2020
- FDAAtellica IM Humidity Pack (Qty 5), SMN 11313505, UDI 00630414234526, Software Version V1.21.0 and lower -March 11, 2020
- FDABard Biopsy EnCor Probe, MRI, 10G Trocar Tip, REF number ECPMR0110G, packaged individually in sterile pouches,March 11, 2020
Questions about this recall
Was RespVent Closed Suction System T-Piece Tracheostomy Adult, Diameter: 4.6mm (14F), Length: 30.5cm (12in.), MDI Adapter, Cont. 10 Closed Suction Catheter, Sterile R, RxOnly, Reorder Number: RTG-02300, UDI: 00850108006035 - Product Usage: To be used to aspirate liquids or semi solids from a patient s upper airway. recalled?
Yes. FDA published recall Z-1443-2020 on March 11, 2020. Reported hazard: It has been determined that some units of catheters were not tightly connected to the thumb port. This may result in the catheter disconnecting from the thumb port which could reduce the suction performance and make it difficult to remove from the patient's trachea.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit