Cobas c8000 Modular Analyzer Series Cobas c502, Part Number 5964067001, a clinical chemistry analyzer intended for the in vitro quantitative / qualitative determination of analytes in body fluids.

Published April 16, 2014 · Recall initiated: March 13, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Initial Cassette Volume Check (ICVC) is not activated on the cobas c 502 module of the cobas 8000 modular analyzer series when manually filled cobas c packs are loaded. This issue occurs with tests that use cobas c pack MULTI or empty pre-labeled cobas c packs. The ICVC feature allows the reagent probe to dive into the reagent cassette and confirm the reagent level in the bottles. This ensures

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Roche Diagnostics Operations, Inc.
Distributor
USA Nationwide Distribution including San Juan PR
Lot
not applicable
Risk classification
Class II
Area (FDA)
device
Country
United States