Pipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
Published April 23, 2014 · Recall initiated: April 1, 2014
Summary
FDA records medical device Class I recall Z-1453-2014, published April 23, 2014 and initiated April 1, 2014. Recalling firm: Micro Therapeutics Inc, Dba Ev3 Neurovascular; Status: Terminated; Reported quantity: 32 units total (23 units in US); Lot: PED Model/ Lot FA7735014 / 9879718 FA7742512 / 9864827 FA7150035 / 9855921 FA7735016 / 9856429 FA7742518 / 9856431 FA7730010 / 9869991 FA7737514 / 9855933 FA7742518 / 9860762 FA7732514 / 9873854 FA7737516 / 9858010 FA7745014 / 9874111 FA7735014 / 9855265 FA7740016 / 9859214 FA7745016 / 9879736 FA7735014 / 9875267 FA7740016 / 9879180 FA7747514 / 9868888. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for the PTFE (polytetrafluoroethylene) coating to delaminate and detach from the delivery wire.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Micro Therapeutics Inc, Dba Ev3 Neurovascular
- Distributor
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Germany, France, UK, Turkey, Argentina, Canada, Colombia, Mexico, , Austria, Belgium, Denmark, Hungary, Italy, Netherlands, Norway, Poland, Portugal, Russia, Slovakia, Slovenia, Spain, Switzerland, and UAE.
- Lot
- PED Model/ Lot FA7735014 / 9879718 FA7742512 / 9864827 FA7150035 / 9855921 FA7735016 / 9856429 FA7742518 / 9856431 FA7730010 / 9869991 FA7737514 / 9855933 FA7742518 / 9860762 FA7732514 / 9873854 FA7737516 / 9858010 FA7745014 / 9874111 FA7735014 / 9855265 FA7740016 / 9859214 FA7745016 / 9879736 FA7735014 / 9875267 FA7740016 / 9879180 FA7747514 / 9868888
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAev3 Pipeline Flex Embolization Device. For neurological endovascular use.April 1, 2020
- FDAAlligator Retrieval Device (ARD). Used in the peripheral and neurovasculature for foreign body removal.April 23, 2014
- FDASpaceLabs Arkon Anesthesia Delivery System with Version 2.0 Software installed. This system includes aApril 23, 2014
- FDASandia Fruit Bar,4 oz.,; Helados La Tapatia Fresno, CA 93706April 23, 2014
- FDAPipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascularSeptember 4, 2013
- FDAMarathon Flow Directed Micro Catheter, Model #105-5055. Product Usage: The Marathon Flow Directed MicroOctober 3, 2012
Questions about this recall
Was Pipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery. recalled?
Yes. FDA published recall Z-1453-2014 on April 23, 2014. Reported hazard: Potential for the PTFE (polytetrafluoroethylene) coating to delaminate and detach from the delivery wire.
What should I do if I have this product?
Voluntary: Firm initiated · Letter