Part510447 Disposable Infusion Pump, ...Elastomeric Pump 600 mL, 2mL/hr/side,... found in the following kits GoPump Kit 510462-BP, 5 " catheter & BIOPATCH GoPump Kit 510558-BP 10 " catheter& BIOPATCH GoPump Kit 64310 Disposable Single Infusion Pump. Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication.
Published June 12, 2013 · Recall initiated: May 10, 2013
Summary
FDA records medical device Class I recall Z-1456-2013, published June 12, 2013 and initiated May 10, 2013. Recalling firm: Symbios Medical Products, LLC; Status: Terminated; Reported quantity: 34517 for all kits; Lot: GoPump Kit 510462-BP, Lot 11-100694- Lot 13-100092 GoPump Kit 510558-BP , Lot 12-100230 - Lot12-101343 GoPump Kit 510643, Lot 12-100332 - Lot 13-100183. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Symbios Medical Products, LLC
- Distributor
- US Nationwide distribution.
- Lot
- GoPump Kit 510462-BP, Lot 11-100694- Lot 13-100092 GoPump Kit 510558-BP , Lot 12-100230 - Lot12-101343 GoPump Kit 510643, Lot 12-100332 - Lot 13-100183
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPart 510449 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 8mL/hr ),... found in the following kitsJune 12, 2013
- FDAPart 510636 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 300 ml perside), 2 to 10 ml/hr. DeliveryJune 12, 2013
- FDAPart 510180 Disposable Single Infusion Pump, ... Elastomeric Pump 300 mL, 2mL/hr,... found in : GoPump KitJune 12, 2013
- FDAPart 510490 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 300 ml per side)( 6mL/hr ),... found in theJune 12, 2013
- FDAPart 510121 Disposable Infusion Pump, ...Elastomeric Pump 300 mL, 4mL/hr,...sterile.. found in the followingJune 12, 2013
- FDAPart 510450 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 10mL/hr ),... found in the following kitsJune 12, 2013
Questions about this recall
Was Part510447 Disposable Infusion Pump, ...Elastomeric Pump 600 mL, 2mL/hr/side,... found in the following kits GoPump Kit 510462-BP, 5 " catheter & BIOPATCH GoPump Kit 510558-BP 10 " catheter& BIOPATCH GoPump Kit 64310 Disposable Single Infusion Pump. Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication. recalled?
Yes. FDA published recall Z-1456-2013 on June 12, 2013. Reported hazard: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the
What should I do if I have this product?
Voluntary: Firm initiated · Letter