Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
Published March 4, 2026 · Recall initiated: January 8, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Medline Industries, LP
- Distributor
- US Nationwide distribution.
- Lot
- UDI-DI (ea) 10653160373582 UDI-DI (case) 00653160373585 Kit lots 2024051380 2024052080 2024052180 2024052280 2024071580 2025061790 2025072990
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States