Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746

Published March 4, 2026 · Recall initiated: January 8, 2026

Summary

FDA records medical device Class II recall Z-1464-2026, published March 4, 2026 and initiated January 8, 2026. Recalling firm: Medline Industries, LP; Status: Ongoing; Reported quantity: 2020; Lot: UDI-DI (ea) 10653160373582 UDI-DI (case) 00653160373585 Kit lots 2024051380 2024052080 2024052180 2024052280 2024071580 2025061790 2025072990. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medline Industries, LP
Distributor
US Nationwide distribution.
Lot
UDI-DI (ea) 10653160373582 UDI-DI (case) 00653160373585 Kit lots 2024051380 2024052080 2024052180 2024052280 2024071580 2025061790 2025072990
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746 recalled?

Yes. FDA published recall Z-1464-2026 on March 4, 2026. Reported hazard: Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.