Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167
Published March 4, 2026 · Recall initiated: January 8, 2026
Summary
FDA records medical device Class II recall Z-1468-2026, published March 4, 2026 and initiated January 8, 2026. Recalling firm: Medline Industries, LP; Status: Ongoing; Reported quantity: 150; Lot: UDI-DI (ea) 10195327015954 UDI-DI (case) 40195327015955 Kit lots 24IME953 25CMA708. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medline Industries, LP
- Distributor
- US Nationwide distribution.
- Lot
- UDI-DI (ea) 10195327015954 UDI-DI (case) 40195327015955 Kit lots 24IME953 25CMA708
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMEDLINE medical convenience kits labeled as: 1) HIP SCOPE PACK, Medline Kit Number/SKU DYNJ64103C; 2) HIPOctober 28, 2026
- FDAMEDLINE medical convenience kits labeled as: 1) BIRTHDAY KIT-DOUBLE BASIN-LF, Medline Kit Number/SKUOctober 28, 2026
- FDAMEDLINE DRAPE C-SECTION FEN CLR SCREEN W, Medline Product Number/SKU DYNJP6108October 28, 2026
- FDAMEDLINE medical convenience kits labeled as: FETOSCOPY CDS-LF, Medline Product Number/SKU CDS840215POctober 28, 2026
- FDAMetoprolol Succinate Extended-Release Tablets, 25 mg, 100 tablets bottles (NDC: 45963-709-11) and 1,000March 11, 2026
- FDAMedline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746March 4, 2026
Questions about this recall
Was Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167 recalled?
Yes. FDA published recall Z-1468-2026 on March 4, 2026. Reported hazard: Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
What should I do if I have this product?
Voluntary: Firm initiated · Letter