FDAFDA · Food and drugs·Recall number Z-1481-2018
Fabius MRI, anesthesia machine, catalog no. 8607300 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). This device can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less.
Published May 16, 2018 · Recall initiated: March 1, 2018
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Draeger Medical, Inc.
- Distributor
- US Nationwide Distribution to accounts in 21 states: AZ, CA, CO, ID, IL, KS, MA, MI, MS, NC, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI; and PR.
- Lot
- Serial no: ASKM-0020 ASKM-0011 ASKL-0005 ASKL-0007 ASKL-0010 ASKM-0031 ASKM-0032 ASKM-0023
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States