Abbott Acclaim Infusion Pump Designed to deliver parenteral infusions including whole blood or red blood cell components and enteral fluids using a wide variety of standard administration sets and fluid containers.

Published May 7, 2014 · Recall initiated: October 31, 2013

Summary

FDA records medical device Class I recall Z-1482-2014, published May 7, 2014 and initiated October 31, 2013. Recalling firm: Hospira Inc.; Status: Terminated; Reported quantity: 13,218 infusion pumps total; Lot: List Number: 12032; Serial Numbers: 12322921, 12323859, 12322780, 12322801, 12322810, 12322812, 12322878, 12322880, 12322887, 12322896, 12322903, 12322904, 12322911, 12322917, 12322936, 12322939, 12323867, 12322916, 12323056, 12321139, 12324767, 12324854, 12324802, 12322873, 12324375, 1231453, 1231610. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Hospira has received customer reports of broken door assemblies on the Hospira/Abbott Acclaim Encore infusion pumps. If the door assembly breaks, it may prevent the door from closing properly and unrestricted flow may occur. If the door cannot be closed, the pump cannot be used which can result in a delay in therapy.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Hospira Inc.
Distributor
Nationwide Distribution -- AK, AL, AR, AZ, CA, CO, CR, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.
Lot
List Number: 12032; Serial Numbers: 12322921, 12323859, 12322780, 12322801, 12322810, 12322812, 12322878, 12322880, 12322887, 12322896, 12322903, 12322904, 12322911, 12322917, 12322936, 12322939, 12323867, 12322916, 12323056, 12321139, 12324767, 12324854, 12324802, 12322873, 12324375, 1231453, 1231610.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Abbott Acclaim Infusion Pump Designed to deliver parenteral infusions including whole blood or red blood cell components and enteral fluids using a wide variety of standard administration sets and fluid containers. recalled?

Yes. FDA published recall Z-1482-2014 on May 7, 2014. Reported hazard: Hospira has received customer reports of broken door assemblies on the Hospira/Abbott Acclaim Encore infusion pumps. If the door assembly breaks, it may prevent the door from closing properly and unrestricted flow may occur. If the door cannot be closed, the pump cannot be used which can result in a delay in therapy.

What should I do if I have this product?

Voluntary: Firm initiated · Letter