Zimmer Air Dermatome II, REF 00-8851-001-00 handpiece Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities

Published June 12, 2013 · Recall initiated: May 7, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.

What to do

Voluntary: Firm initiated · Telephone

Recall details

Manufacturer
Zimmer Surgical Inc
Distributor
Worldwide distribution: USA Nationwide including Puerto Rico; UAE, TAIWAN, SWEDEN, THAILAND, KOREA, SINGAPORE, HONG KONG CHINA, CHILE, CANADA, AUSTRALIA, JAPAN, ENGLAND, GERMANY, and EGYPT.
Lot
Serial Number 500191 , 500192,500193, 500194,500196,500197,500198,500199, 500200, 500201,500202,500203, 500204, 500205,500206, 500207,500208,500209,500210, 500211,500212, 500213,500215,500216,500217, 500219,500220, 500221,500222, 500223, 500224, 500225,500226,500227,500228,500229,500230, 500231,500232, 500233,500234,500235,500236, 500237, 500238,500239,500240,500241 , 500242, 500243,500244,500245,500246,500247,500248,500249, 500250, 500251,500252,500253,500254, 500255,500256,500257,500258,500259, 500260,500261,500262,500263,500264, 500265,500266,500267,500268,500269, 500270,500291, 500292,500293,500295,500296,500297,500298, 500299,500300,500301 , 500302,500303, 500304, 500306, 500307,500308,500309,500310, 500331 , 500332, 500334, 500335, 500336, 500337, 500338, 500339, 500340, 500341,500342, 500343, 500344, 500345, 500346, 500347, 500348,500349, 500350, 500351 , 500352, 500353, 500354,500355, 500357, 500358, 500359, 500368, 500369,500370,500371
Risk classification
Class II