FDAFDA · Food and drugs·Recall number Z-1492-2017
Brilliance BigBore Oncology CT - 728243 Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories. This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.
Published March 29, 2017 · Recall initiated: February 23, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%).
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Worldwide Distribution - US to AR, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MO, NC, NE, NJ, NM, NY, OH, TX & WI; Internationally to the following countries: Australia, Austria, Canada, China, Czech Republic, Germany, India, Israel, Jordan, Korea, Mexico, New Cldonia, Romania, Russian Federation, Turkey, & United Kingdom.
- Lot
- System Serial Number - 76002, 76008, 76010, 76011, 76016, 76020, 76021, 76022, 76023, 76024, 76025
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States