Stryker PenAdapt"; Rx Only, Sterile. This device is an accessory to an electrosurgical unit (ESU). The PenAdapt is used to remove smoke, particles, and body and casual fluids from the point of surgical activity during surgical procedures that use ESU for cutting and cauterizing.
Published April 30, 2014 · Recall initiated: March 26, 2014
Summary
FDA records medical device Class II recall Z-1493-2014, published April 30, 2014 and initiated March 26, 2014. Recalling firm: Stryker Instruments Div. of Stryker Corporation; Status: Terminated; Reported quantity: 35,310 unexpired units; Lot: Part Number 0702-045-027. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During packaging verification testing, a failure occurred due to cuts in the packaging potentially causing a breach in the sterile barrier. Additional testing identified microscopic aluminum metal shavings and foreign particles estimated in size to be between 0.01inches x 0.015 inches to 0.03 inches x 0.04 inches that were introduced during the manufacturing process and were sterilized within the
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Instruments Div. of Stryker Corporation
- Distributor
- Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV,and WY, and the countries of Austria, Canada, Sweden, England, UAE, Netherlands, Spain, and Hong Kong.
- Lot
- Part Number 0702-045-027
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting,July 27, 2022
- FDANeptune E-SEP 165mm Blade Electrode, Catalog Number 0703-165-001 - Product Usage: The Neptune E-SEP is anMay 6, 2020
- FDAColor Cuff Disposable Tourniquet Cuff, 18 inch X 3 inch, 1BLA (Single Bladder), 1PRT(Single Port) QuickDecember 11, 2019
- FDAC2 Nerve Monitor, 8 Channel, 5140-508-280, UDI 07613327277319 Product Usage: The C2 Nerve Monitor System isOctober 9, 2019
- FDAProteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116.April 30, 2014
- FDAArrow Pressure Injectable PICC Powered by Arrow VPS Stylet Catalogue number CDC-35052-VPS, peripherallyApril 30, 2014
Questions about this recall
Was Stryker PenAdapt"; Rx Only, Sterile. This device is an accessory to an electrosurgical unit (ESU). The PenAdapt is used to remove smoke, particles, and body and casual fluids from the point of surgical activity during surgical procedures that use ESU for cutting and cauterizing. recalled?
Yes. FDA published recall Z-1493-2014 on April 30, 2014. Reported hazard: During packaging verification testing, a failure occurred due to cuts in the packaging potentially causing a breach in the sterile barrier. Additional testing identified microscopic aluminum metal shavings and foreign particles estimated in size to be between 0.01inches x 0.015 inches to 0.03 inches x 0.04 inches that were introduced during the manufacturing process and were sterilized within the
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit