Cardio Vascular-Allura Centron, system code 722400

Published March 25, 2020 · Recall initiated: December 30, 2019

Summary

FDA records medical device Class II recall Z-1494-2020, published March 25, 2020 and initiated December 30, 2019. Recalling firm: Philips North America, LLC; Status: Terminated; Reported quantity: 87; Lot: Not distributed in US. Serial numbers 132 133 49 66 71 135 31 36 79 69 137 149 147 109 136 108 103 93 53 127 129 129 55 74 2 57 106 56 86 68 157 78 64 63 23 51 100 97 102 113 90 44 30 101 77 83 67 151 22 70 96 116 131 35 82 9 91 43 61 161 28 34 27 19 144 92 168 172 173 122 197 139 6 76 176 16 130 60 80 184 112 81 89 42 110 155 33. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips North America, LLC
Distributor
Nationwide domestic distribution. Foreign distribution worldwide.
Lot
Not distributed in US. Serial numbers 132 133 49 66 71 135 31 36 79 69 137 149 147 109 136 108 103 93 53 127 129 129 55 74 2 57 106 56 86 68 157 78 64 63 23 51 100 97 102 113 90 44 30 101 77 83 67 151 22 70 96 116 131 35 82 9 91 43 61 161 28 34 27 19 144 92 168 172 173 122 197 139 6 76 176 16 130 60 80 184 112 81 89 42 110 155 33
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Cardio Vascular-Allura Centron, system code 722400 recalled?

Yes. FDA published recall Z-1494-2020 on March 25, 2020. Reported hazard: A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.

What should I do if I have this product?

Voluntary: Firm initiated · Letter