Cardio Vascular-Allura Centron, system code 722400
Published March 25, 2020 · Recall initiated: December 30, 2019
Summary
FDA records medical device Class II recall Z-1494-2020, published March 25, 2020 and initiated December 30, 2019. Recalling firm: Philips North America, LLC; Status: Terminated; Reported quantity: 87; Lot: Not distributed in US. Serial numbers 132 133 49 66 71 135 31 36 79 69 137 149 147 109 136 108 103 93 53 127 129 129 55 74 2 57 106 56 86 68 157 78 64 63 23 51 100 97 102 113 90 44 30 101 77 83 67 151 22 70 96 116 131 35 82 9 91 43 61 161 28 34 27 19 144 92 168 172 173 122 197 139 6 76 176 16 130 60 80 184 112 81 89 42 110 155 33. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America, LLC
- Distributor
- Nationwide domestic distribution. Foreign distribution worldwide.
- Lot
- Not distributed in US. Serial numbers 132 133 49 66 71 135 31 36 79 69 137 149 147 109 136 108 103 93 53 127 129 129 55 74 2 57 106 56 86 68 157 78 64 63 23 51 100 97 102 113 90 44 30 101 77 83 67 151 22 70 96 116 131 35 82 9 91 43 61 161 28 34 27 19 144 92 168 172 173 122 197 139 6 76 176 16 130 60 80 184 112 81 89 42 110 155 33
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe IFU for the Autoclavable Temperature Probes: Part # 453564635891, providing in the packaging for theApril 20, 2022
- FDAPhilips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A,November 4, 2020
- FDAPhilips SureSigns VM1SpO2, C02, Rec, model no. 863266 - Product Usage: The SureSigns VM1 is a patient monitorJune 17, 2020
- FDAPhilips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospitalMay 27, 2020
- FDAUroDiagnost Eleva, system code 708033March 25, 2020
- FDAArtis Q Biplane, Model No. 10848282; Artis Q Ceiling, Model No.10848281; Artis Q Floor, Model No. 10848280;March 25, 2020
Questions about this recall
Was Cardio Vascular-Allura Centron, system code 722400 recalled?
Yes. FDA published recall Z-1494-2020 on March 25, 2020. Reported hazard: A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
What should I do if I have this product?
Voluntary: Firm initiated · Letter