Merge Cardio software with Issuer of Patient ID (IPID). The firm name on the label is Merge Healthcare, Hartland, WI.
Published March 29, 2017 · Recall initiated: September 22, 2015
Summary
FDA records medical device Class II recall Z-1496-2017, published March 29, 2017 and initiated September 22, 2015. Recalling firm: Merge Healthcare, Inc.; Status: Terminated; Reported quantity: 16 sites were originally identified as potentially having the affected versions; Lot: Versions 8.30, 9.0, 9.0.1, 9.0.2, 9.0.3, 9.0.4, 9.0.5, 9.0.6, 9.0.7, 9.0.8, 10.0, 10.0.1, and 10.1 with Issuer of Patient ID (IPID). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
For sites using the Issuer of Patient ID (IPID), the system will display the study list and images from different patients with different IPIDs as though they are for the same patient if they all have the first name, last name, and medical record number in common.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Merge Healthcare, Inc.
- Distributor
- US Distribution was made to medical facilities located in GA, IL, MD, MI, MO, OH, TN, and VT. There was no foreign/government/military distribution.
- Lot
- Versions 8.30, 9.0, 9.0.1, 9.0.2, 9.0.3, 9.0.4, 9.0.5, 9.0.6, 9.0.7, 9.0.8, 10.0, 10.0.1, and 10.1 with Issuer of Patient ID (IPID)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMerge PACS and Merge OrthoPACS, medical imaging diagnostic workstationSeptember 2, 2026
- FDAMerge OrthoPACS, medical imaging diagnostic workstationAugust 12, 2026
- FDAMerge PACS, medical imaging diagnostic workstationAugust 12, 2026
- FDAVERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is anMarch 26, 2025
- FDAOrthosize Templating Version 1.2.6 Echo Bi-Metric Hip Stem Digital Templates. For preoperative planning ofMarch 29, 2017
- FDAMerge Eye Station Import Utility (ESIU). The firm name on the labeling is Merge Healthcare.March 29, 2017
Questions about this recall
Was Merge Cardio software with Issuer of Patient ID (IPID). The firm name on the label is Merge Healthcare, Hartland, WI. recalled?
Yes. FDA published recall Z-1496-2017 on March 29, 2017. Reported hazard: For sites using the Issuer of Patient ID (IPID), the system will display the study list and images from different patients with different IPIDs as though they are for the same patient if they all have the first name, last name, and medical record number in common.
What should I do if I have this product?
Voluntary: Firm initiated · Letter