Back to search
FDAFDA · Food and drugs·Recall number Z-1503-2014

Digital Diagnost, stationary radiographic system; Models: 712020/22/25/26/50/52/54/55/57/60/62/80/81/82/85

Published April 30, 2014 · Recall initiated: March 17, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips Medical Systems, Inc.
Distributor
Nationwide: US: AL, AR, AZ, CA, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, WY / Worldwide including Canada & Mexico
Lot
Units distributed from November 2006 through March 2014.
Risk classification
Class II
Area (FDA)
device
Country
United States