Digital Diagnost, stationary radiographic system; Models: 712020/22/25/26/50/52/54/55/57/60/62/80/81/82/85
Published April 30, 2014 · Recall initiated: March 17, 2014
Summary
FDA records medical device Class II recall Z-1503-2014, published April 30, 2014 and initiated March 17, 2014. Recalling firm: Philips Medical Systems, Inc.; Status: Terminated; Reported quantity: 985 devices; Lot: Units distributed from November 2006 through March 2014. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems, Inc.
- Distributor
- Nationwide: US: AL, AR, AZ, CA, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, WY / Worldwide including Canada & Mexico
- Lot
- Units distributed from November 2006 through March 2014.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFoot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 ThisApril 6, 2016
- FDAPhilips Healthcare Xper vascular systems R7.6 Model : 722134 Product Usage: The Allura Xper series is intendedDecember 9, 2015
- FDAPhilips Healthcare INTEGRIS vascular System Model : 722122 Product Usage: The Allura Xper series is intendedDecember 9, 2015
- FDAPhilips Healthcare INTEGRIS H5000F/Allura 9F Model : 722017 Product Usage: The Allura Xper series is intendedDecember 9, 2015
- FDAOn-Board Imager - Exact Arm Elbow Motor Shaft (kVD, kVS, and MV) with ''Type 01" elbow motors. Product Usage:April 30, 2014
- FDASiemens ACUSON X700 Ultrasound Systems with software version 1.0.00, 1.0.01, 1.0.02, 1.1.00 and 1.1.02.April 30, 2014
Questions about this recall
Was Digital Diagnost, stationary radiographic system; Models: 712020/22/25/26/50/52/54/55/57/60/62/80/81/82/85 recalled?
Yes. FDA published recall Z-1503-2014 on April 30, 2014. Reported hazard: The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held.
What should I do if I have this product?
Voluntary: Firm initiated · Letter