FDAFDA · Food and drugs·Recall number Z-1513-2014

ExacTrac 6.0 is a patient positioning and monitoring system. Model/catalogue numbers: 20833B EXACTRAC 6.0 IR POSITIONING SOFTWARE 49936 ET SOFTWARE UPDATE 6.0.X TO 6.0.3 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED 49973B ET DATA PREP / REVIEW SYSTEM 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED

Published May 7, 2014 · Recall initiated: February 24, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point. As a result, there is a possibility that the radiation treatment dose may be delivered to the unintended target position.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Brainlab AG
Distributor
Worldwide Distribution -- USA, including the states of CA, IN, LA, MA, OH, TX, and WI, and the countries of Australia, China, France, Germany, Japan, Netherlands, and the United Kingdom.
Lot
ExacTrac version v.6.0.3.
Risk classification
Class II
Area (FDA)
device
Country
Germany