FDAFDA · Food and drugs·Recall number Z-1515-2024

EndoVive 20Fr Push Safety PEG Kit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M00566471

Published April 17, 2024 · Recall initiated: February 20, 2024

Summary

FDA records medical device Class II recall Z-1515-2024, published April 17, 2024 and initiated February 20, 2024. Recalling firm: Boston Scientific Corporation; Status: Ongoing; Reported quantity: 2501 units; Lot: GTIN: 08714729802631 Lot Numbers: 32295181 , 32295182 , 32306054 , 32306055 , 32316800 , 32323673 , 32342013 ,32367049 , 32367050 , 32374782 , 32472868 , 32486230 , 32486231 , 32486232 , 32486233 , 32506056 , 32515039 , 32523243 , 32523244 , 32531793 , 32531794 , 32531795 , 33113026 , 33113027 , 33113028 , 33113029 , 33170338. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Boston Scientific Corporation
Distributor
Worldwide distribution - US Nationwide and the country of Japan.
Lot
GTIN: 08714729802631 Lot Numbers: 32295181 , 32295182 , 32306054 , 32306055 , 32316800 , 32323673 , 32342013 ,32367049 , 32367050 , 32374782 , 32472868 , 32486230 , 32486231 , 32486232 , 32486233 , 32506056 , 32515039 , 32523243 , 32523244 , 32531793 , 32531794 , 32531795 , 33113026 , 33113027 , 33113028 , 33113029 , 33170338
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was EndoVive 20Fr Push Safety PEG Kit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M00566471 recalled?

Yes. FDA published recall Z-1515-2024 on April 17, 2024. Reported hazard: Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration

What should I do if I have this product?

Voluntary: Firm initiated · Letter