Siemens SOMATOM Definition: Intended to produce cross-sectional images of the body. Computed tomography (CT) x-ray system.

Published April 27, 2016 · Recall initiated: March 2, 2016

Summary

FDA records medical device Class II recall Z-1518-2016, published April 27, 2016 and initiated March 2, 2016. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 1096; Lot: Model Number: 7740769 SOMATOM Definition family of CT systems running on software versions VA44A_SP3a, VA44A_SP3b, VA44A_SP3c. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software bugs in these SOMATOM systems could possibly cause scan aborts, re-scans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium, accidental x-radiation to the user or impede customer workflow.

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
Nationwide Distribution-including DC, and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
Lot
Model Number: 7740769 SOMATOM Definition family of CT systems running on software versions VA44A_SP3a, VA44A_SP3b, VA44A_SP3c.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Siemens SOMATOM Definition: Intended to produce cross-sectional images of the body. Computed tomography (CT) x-ray system. recalled?

Yes. FDA published recall Z-1518-2016 on April 27, 2016. Reported hazard: Software bugs in these SOMATOM systems could possibly cause scan aborts, re-scans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium, accidental x-radiation to the user or impede customer workflow.

What should I do if I have this product?

Voluntary: Firm initiated