INOmax DSIR, model 10007, Nitric Oxide Delivery System, with Software version 3.0.0. The INOmax DS delivery system delivers INOMAX (nitric oxide for inhalation) therapy gas and provides continuous monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Published May 14, 2014 · Recall initiated: March 27, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

An anomaly impacts the monitoring system of INOmax DSIR Plus (software 3.0.0). When the integrated pneumatic backup switch is activated with low or no gas flow through the injector module, an intended NO cell failure alarm will occur. In addition, an unintended NO2 cell failure alarm may also occur. The NO/NO2 cell failure alarms are caused by an average NO/NO2 circuit concentration greater tha

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
INO Therapeutics (dba Ikaria)
Distributor
Nationwide Distribution including DC and VA.
Lot
Serial Numbers: DS20080152, DS20090218, DS20090243, DS20090271, DS20090523, DS20090598, DS20090821, DS20090925, DS20101022, DS20101076, DS20101147, DS20101374, DS20101413, DS20101420, DS20101499, DS20101637, DS20111199, DS20120099, DS20120150, DS20120169, DS20120288, DS20120306.
Risk classification
Class II
Area (FDA)
device
Country
United States