FDAFDA · Food and drugs·Recall number Z-1540-2016

TC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 2 RIGHT, STERILE R, REF 12 000 009, S&N 7500264, Smith & Nephew Orthopaedics AG Product Usage: Tibial knee prosthesis

Published April 27, 2016 · Recall initiated: March 11, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being unable to seat the polyethylene (PE) Insert on the affected VKS tibial base plate during surgery.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Smith & Nephew, Inc.
Distributor
International Distribution only in countries of: Germany and Switzerland.
Lot
Batch number C1507853
Risk classification
Class II
Area (FDA)
device
Country
United States