Back to search
FDAFDA · Food and drugs·Recall number Z-1542-2024

Smiths Medical Model 3010 Syringe pumps: a. Protege Model 3010, b. Medfusion Model 3010A, c. Protege UK Model 3010E

Published April 24, 2024 · Recall initiated: December 19, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Smiths Medical ASD Inc.
Distributor
Worldwide distribution.
Lot
All serial numbers for pumps manufactured or serviced with these potentially affected Barrel Clamp Guides between Jul-2016 and Apr-2021. a. 3010, UDI/DI none, Serial Numbers: M03315, M03910; b. 3010A, UDI/DI 10610586032097, Serial Numbers: M04601, M05551, M05782, M06829, M07259, M07334, M07339, M07342, M07345, M07371, M08269, M08750, M08815, M10253, M10267, M10871, M10876, M10965, M11872, M11896, M11899, M11901, M11905, M11915; c. 3010E, UDI/DI none, Serial Numbers: M05179
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card