Biomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma Fixation Systems. Product Usage: The Phoenix Retrograde Femoral Nail is indicated for alignment, stabilization, and fixation of fractures caused by trauma or disease and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity.
Published June 26, 2013 · Recall initiated: April 29, 2013
Summary
FDA records medical device Class II recall Z-1547-2013, published June 26, 2013 and initiated April 29, 2013. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 75; Lot: Catalog number: 14-442021 and lot number:757910. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Biomet Trauma ("Biomet") has initiated a recall of Retrograde Femoral Connecting Bolt, which involves Part Number 14-442021. The connecting bolt has an undersized diameter specification that may cause an interference fit with the 4mm hex driver (Part Number: 41024) near the edge of the tolerance. If the 4mm hex driver becomes stuck in the connecting bolt it cannot be taken apart and a delay in su
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomet, Inc.
- Distributor
- Worldwide Distribution - USA Nationwide: CA, TX, NY, PA, NJ, FL, OH, MI, CO, SD, IN, MT, GA, and WI. and the countries of Netherlands, Costa Rica, Japan, and San Juan.
- Lot
- Catalog number: 14-442021 and lot number:757910
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475640 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475625 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475630 Product Description: Intramedullary BoneAugust 12, 2026
- FDAZimmer Patient Helper Bed Frame Adapter Assembly The bed frame adapter assembly is intended to be used byJune 26, 2013
- FDASuperior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulderJune 26, 2013
Questions about this recall
Was Biomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma Fixation Systems. Product Usage: The Phoenix Retrograde Femoral Nail is indicated for alignment, stabilization, and fixation of fractures caused by trauma or disease and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity. recalled?
Yes. FDA published recall Z-1547-2013 on June 26, 2013. Reported hazard: Biomet Trauma ("Biomet") has initiated a recall of Retrograde Femoral Connecting Bolt, which involves Part Number 14-442021. The connecting bolt has an undersized diameter specification that may cause an interference fit with the 4mm hex driver (Part Number: 41024) near the edge of the tolerance. If the 4mm hex driver becomes stuck in the connecting bolt it cannot be taken apart and a delay in su
What should I do if I have this product?
Voluntary: Firm initiated · Letter